Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023

Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023 : Khóa luận tốt nghiệp Cử nhân kỹ thuật y học

 HP., 2024
 HP. : 59tr. ; 27cm Tiếng Việt
Mô tả biểu ghi
ID:20421
Kí hiệu phân loại B16
Tác giả CN Duong Do Thanh Hang
Nhan đề Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023
Phụ đề Khóa luận tốt nghiệp Cử nhân kỹ thuật y học
Thông tin xuất bản HP. :HP.,2024
Mô tả vật lý 59tr. ;27cm
Tóm tắt Evaluation of method precision revealed that the estimated standard deviation at HPC was smaller than the manufacturer's declared value ( =0.0528 < 0.06= ; =0.0536 < 0.06= ), confirming the precision for HPC. The coefficient of variation (CV%) for LPC was smaller than the UVL value ( =4.55 < 4.75=UVL; =4.55 < 4.75=UVL), hence confirming the precision for LPC. Accuracy assessment showed PPA, NPA, OPA=100%, ̅ SD of LPC and HPC falling within the manufacturer's recommended range across runs. Linearity assessment indicated =0.99, slope=0.92, confirming the method's linearity. 2. HBV DNA load exhibits a weak linear correlation with HBcrAg concentration in the patient group with HBV DNA load above the quantification threshold (r = 0.27, p < 0.05). The correlation between HBV DNA load and HBcrAg concentration in the untreated group (r = 0.59, p < 0.05) is stronger than in the group treated with antiviral drugs (r = 0.48, p < 0.05)
Thuật ngữ chủ đề Kỹ thuật y học
Từ khóa tự do HBcrAg; Kỹ thuật y học; Tải lượng HBV DNA
Tác giả(bs) CN Phạm Thị Thu Trang - hd.
Tác giả(bs) CN Trần Quang Đôn - hd.
Địa chỉ 100Phòng tra cứu(1): PTC03437
MARC
Hiển thị đầy đủ trường & trường con
TagGiá trị
00000000nam#s2200000ua#4500
00120421
0027
004ILIB-23231
005202409040909
008240904s2024 ||||||viesd
0091 0
039[ ] |a 202409040932 |b ILIB |y 202409040931 |z ILIB
041[ ] |a vie
084[ ] |a B16
100[ ] |a Duong Do Thanh Hang
245[ ] |a Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023 : |b Khóa luận tốt nghiệp Cử nhân kỹ thuật y học / |c Duong Do Thanh Hang
260[ ] |a HP. : |b HP., |c 2024
300[ ] |a 59tr. ; |c 27cm |e File
520[ ] |a Evaluation of method precision revealed that the estimated standard deviation at HPC was smaller than the manufacturer's declared value ( =0.0528 < 0.06= ; =0.0536 < 0.06= ), confirming the precision for HPC. The coefficient of variation (CV%) for LPC was smaller than the UVL value ( =4.55 < 4.75=UVL; =4.55 < 4.75=UVL), hence confirming the precision for LPC. Accuracy assessment showed PPA, NPA, OPA=100%, ̅ SD of LPC and HPC falling within the manufacturer's recommended range across runs. Linearity assessment indicated =0.99, slope=0.92, confirming the method's linearity. 2. HBV DNA load exhibits a weak linear correlation with HBcrAg concentration in the patient group with HBV DNA load above the quantification threshold (r = 0.27, p < 0.05). The correlation between HBV DNA load and HBcrAg concentration in the untreated group (r = 0.59, p < 0.05) is stronger than in the group treated with antiviral drugs (r = 0.48, p < 0.05)
650[ ] |a Kỹ thuật y học
653[ ] |a HBcrAg
653[ ] |a Kỹ thuật y học
653[ ] |a Tải lượng HBV DNA
700[ ] |a Phạm Thị Thu Trang |e hd.
700[ ] |a Trần Quang Đôn |e hd.
852[ ] |a 100 |b Phòng tra cứu |j (1): PTC03437
890[ ] |a 1 |b 0 |c 0 |d 0
Dòng Mã vạch Bản sao Nơi lưu Tình trạng Cho phép yêu cầu
1 PTC03437 1 Kho tra cứu
#1 PTC03437
Nơi lưu Kho tra cứu
Tình trạng
Bình luận