Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023

Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023 : Khóa luận tốt nghiệp Cử nhân kỹ thuật y học

 HP., 2024
 HP. : 59tr. ; 27cm Tiếng Việt
Mô tả biểu ghi
ID:20820
Kí hiệu phân loại B16
Tác giả CN Duong Do Thanh Hang
Nhan đề Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023 : Khóa luận tốt nghiệp Cử nhân kỹ thuật y học / Duong Do Thanh Hang
Thông tin xuất bản HP. :HP.,2024
Mô tả vật lý 59tr. ;27cm
Tóm tắt Evaluation of method precision revealed that the estimated standard deviation at HPC was smaller than the manufacturer's declared value ( =0.0528 < 0.06= ; =0.0536 < 0.06= ), confirming the precision for HPC. The coefficient of variation (CV%) for LPC was smaller than the UVL value ( =4.55 < 4.75=UVL; =4.55 < 4.75=UVL), hence confirming the precision for LPC. Accuracy assessment showed PPA, NPA, OPA=100%, ̅ SD of LPC and HPC falling within the manufacturer's recommended range across runs. Linearity assessment indicated =0.99, slope=0.92, confirming the method's linearity. 2. HBV DNA load exhibits a weak linear correlation with HBcrAg concentration in the patient group with HBV DNA load above the quantification threshold (r = 0.27, p < 0.05). The correlation between HBV DNA load and HBcrAg concentration in the untreated group (r = 0.59, p < 0.05) is stronger than in the group treated with antiviral drugs (r = 0.48, p < 0.05)
Thuật ngữ chủ đề Kỹ thuật y học
Từ khóa tự do HBcrAg
Từ khóa tự do Kỹ thuật y học
Từ khóa tự do Tải lượng HBV DNA
Tác giả(bs) CN Phạm Thị Thu Trang
Tác giả(bs) CN Trần Quang Đôn
Địa chỉ 100Phòng tra cứu(1): PTC03437
MARC
Hiển thị đầy đủ trường & trường con
TagGiá trị
00000000naa#s2200000ua#4500
00120820
0027
004ILIB-23231
005202409040909
008240904s2024 ||||||viesd
0091 0
039[ ] |a 20260331200857 |b admin |c 202409040932 |d ILIB |y 202409040931 |z ILIB
041[ ] |a vie
084[ ] |a B16
100[ ] |a Duong Do Thanh Hang
245[ ] |a Verification of the real-time PCR testing method for measuring the HBV DNA load on the Cobas 5800 automated system at Hai Phong International Hospital in 2023 : |b Khóa luận tốt nghiệp Cử nhân kỹ thuật y học / |c Duong Do Thanh Hang
260[ ] |a HP. : |b HP., |c 2024
300[ ] |a 59tr. ; |c 27cm |e File
520[ ] |a Evaluation of method precision revealed that the estimated standard deviation at HPC was smaller than the manufacturer's declared value ( =0.0528 < 0.06= ; =0.0536 < 0.06= ), confirming the precision for HPC. The coefficient of variation (CV%) for LPC was smaller than the UVL value ( =4.55 < 4.75=UVL; =4.55 < 4.75=UVL), hence confirming the precision for LPC. Accuracy assessment showed PPA, NPA, OPA=100%, ̅ SD of LPC and HPC falling within the manufacturer's recommended range across runs. Linearity assessment indicated =0.99, slope=0.92, confirming the method's linearity. 2. HBV DNA load exhibits a weak linear correlation with HBcrAg concentration in the patient group with HBV DNA load above the quantification threshold (r = 0.27, p < 0.05). The correlation between HBV DNA load and HBcrAg concentration in the untreated group (r = 0.59, p < 0.05) is stronger than in the group treated with antiviral drugs (r = 0.48, p < 0.05)
650[ ] |a Kỹ thuật y học
653[ ] |a HBcrAg
653[ ] |a Kỹ thuật y học
653[ ] |a Tải lượng HBV DNA
700[ ] |a Phạm Thị Thu Trang |e hd.
700[ ] |a Trần Quang Đôn |e hd.
852[ ] |a 100 |b Phòng tra cứu |j (1): PTC03437
890[ ] |a 1 |b 0 |c 0 |d 0
Dòng Mã vạch Bản sao Nơi lưu Tình trạng Cho phép yêu cầu
1 PTC03437 1 Phòng tra cứu
#1 PTC03437
Nơi lưu Phòng tra cứu
Tình trạng
Bình luận